The APT 100 is a device for measuring tensile strength of seals made on medical bags. The APT 100 is designed to perform peel tests following the ASTM F88-00
as well as the EN868-5. The machine clamps a specimen and pulls the seal apart, measuring the tensile strength with load cell. The machine collects the data 100 times per second and displays the maximum seal strength.

By transferring the data to a PC, the complete test result will be presented in table and colored graphic chart. Besides the seal strength, the seal width can also be measured according to your own seal strength definition. The result can be printed out as peel test report for one of your validation processes.


Images

 

Video Coming Soon

See More on Our
Youtube Channel


PDF

PDFimage

  • MHS 300
  • Medical sealer
  • Magvac Med
  • IST Med
  • Medical Heat Sealer
  • Medical Rotary Sealer
  • 411 MSIK
  • 611MSIK
  • 411 MSIDK
  • 611 MSIDK
  • Contimed D 662 MD
  • Contimed D 665 MD
  • Contimed D 665 MP
  • D665 MD
  • D665 MD
  • D770 MPCV
  • D775 MPCV
  • APT 100 Peel Tester
  • ATM Temperature Measurement
Audion Medical Sealer

STERILE PACKAGING IN THE

MEDICAL INDUSTRY

Healthcare is one of the most important fundamentals in our

life; mankind is spending lots of money on a healthy lifestyle,

health insurances and new developments, simply because

good health is unaffordable.

Audion supplies several validatable sealers for the medical

industry/healthcare. The machines have a strong price-quality

ratio and are easy to work with. All machines are built according

to the latest directives and create high quality seals. The

different options and functions make Audion medical sealers

suitable for a variety of medical packaging purposes.

In addition to the packaging machines, we can also supply

equipment to support the validation process.

 

 

Specification

Technical features

  • Maximum measuring force : 100,0 N
  • Measuring unit : N, Kgf, Lbf
  • Sampling : 100 data/second
  • Increments : 0,1 N
  • Measuring accuracy : +/- 0,4 % of full scale
  • Data transfer interface : USB cable
  • Operator / Administrator registration
  • Measuring setting memory : 5
  • Test data storage : 120
  • Specimen material thickness range : 30 – 200 micron (per clamping layer)

Measuring settings

  • Peel origin (starting position) : 10 mm or 25 mm
  • Peel speed : 200 mm/min. or 300 mm/min.
  • Test end detection : Auto or Manual

Judgment criteria

  • Max. seal strength upper limit : 0,1 – 100,0 N
  • Max. seal strength lower limit : 0,0 – 99,9 N
  • Effective seal width : 6,0 mm
  • Qualified seal width : 0 – 10,0 mm

Specifications

  • Power supply : AC 100 – 240 V, 1 phase, 50/60 Hz
  • Power consumption : 9 W
  • Machine size (LxWxH) : 370 x 270 x 200 mm